Table of Contents Symbol Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 System Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Set-up And Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Lamp Replacement .
Symbol Descriptions Attention: Read Operating Manual for Warnings, Precautions, and Instructions for Use. Dangerous Voltage: Refer Servicing to Qualified Personnel. Type B Equipment Caution — Hot Warning: Risk of Fire. Replace Fuses as Marked. Light; Lighting; Illumination Intensity Indications The Insight Instruments Light Source is specifically designed for use in ophthalmic procedures with the Insight Instruments Mini Light System MLS (P/N: 11003.000) and Wide Angle Light Pipe WALP (P/N: 11001.
WARNING: possible explosion hazard. Do not use in the presence of flammable anesthetics. To ensure patient and operator safety, read and understand this manual before using the Insight Instruments Light Source. Do not remove cover. Electrical shock hazard due to high internal voltage. There are no user serviceable parts inside except for the lamp and fuse, which are accessible through the bottom and back cover panels. Refer all service to Insight Instruments. Replace fuses as marked.
CAUTION: Only qualified medical personnel should use the Light Source. Provide adequate ventilation to prevent overheating. Do not cover or drape this Light Source. Provide a 3 inch (7.62 cm) distance between the Insight Instruments Light Source and any solid objects. Grounding reliability is achieved only when connected to hospital-use or hospital-grade receptacles. Inspect electrical plug and cord routinely. Do not use if damage is discovered.
System Description The Insight Instruments Light Source is used to supply high quality, white light for video and optical illumination systems. It accepts Insight Instruments ACMI connectors.
Set-up And Operation After a 15-second warm-up period, the light output will stabilize at its normal intensity. Plug in the fiber optic bundle to stabilize light output. CAUTION: Verify that the supply voltage matches the voltage range indicated on the rear of the Insight Instruments Light Source. To prevent temporary blinding, turn power on only after the fiber optic bundle is plugged in and the intensity control knob is set to the minimum position.
Mounting Secure bracket to a flat surface with screws provided. Align mounting post on Light Source with mounting hole on bracket (View A). Slide mounting lock until it clicks into place (View B). Light Source is now securely mounted to table. CAUTION: Do not cover or drape, Light Source generates heat. For proper ventilation, maintain a clearance of 3 inches (7.62 cm) around the unit, especially around the fan.
Lamp Replacement CAUTION: Lamp is hot. Allow the lamp to cool before replacing. 1. Turn unit off and unplug the power cord from both the wall outlet and the Insight Instruments Light Source. 2. Remove the lamp door from the bottom of the Insight Instruments Light Source by unscrewing the lamp access knob on the back of the lightbox. Lamp Access Door ts Instrumen t Insigh Lamp Access Knob Instruments Insight Insig ht ts Instrumen Insight Instruments 3.
5. Reconnect the lamp to the lamp connector and place in the spring mechanism. Verify that the pin on the lamp is properly seated in the slot on the lamp bracket. 6. Reattach the lamp door securely. The unit will not power up unless the lamp door is seated properly. Fuse Replacement 1. Turn the unit off and remove the power cord from both the wall outlet and the Insight Instruments Light Source. Fuse Holder Fuse Holder 2.
5. Reinsert the fuse holder by pressing it into the power connector until it snaps into place. 6. If the new fuse blows, immediately contact Insight Instruments Technical Service at 1-407-324-2727. Cleaning NOTE: Do not sterilize or immerse the Insight Instruments Light Source. 1. Prior to cleaning, turn power off and disconnect the power cord from both the wall outlet and the Insight Instruments Light Source. 2.
Specifications Electrical Input Voltage: 100 - 240 VAC, 50 - 60 Hz Input Current: 1A (@ 100 VAC) Output: 60 VDC @ 0.4A Leakage Current: less than 100 microamps Dimensions 8.2" (21 cm) (L) x 3.1" (8 cm) (W) x 5.2" (13 cm) (H) Illumination System Insight Instruments Solarc™ Lamp Color Temperature: 5500°K Lamp: 60 Volts, 24 watts Lamp Life: 300 hours average Dimming System Manual Shutter: 20% to 100% Fiber Optic Bundle Interface Insight Instruments ACMI Connector Approvals UL 2601-1 CSA C22.2 NO 601.
Equipment Classification Class I, Type B, Continuous Operation Fuses T1.0A L250V: 1.
Troubleshooting If lamp is not illuminating and... fan is not operating. 12 POSSIBLE CAUSE SOLUTION Light Source is not Plug the power cord into the plugged into either the wall outlet and power cord wall outlet or the socket at back of unit. power cord socket. Power switch is not turned on. Depress the “I” side of the power switch. The bottom cover is not properly installed.
fan is operating. The lamp was not able Wait at least 15 seconds after to “cool down” before turning the Light Source off restarting. before turning it back on. The Light Source has overheated. Verify that all ventilation slots and the fan area are not obstructed. Mount the Light Source on the supplied mounting bracket and verify there is always at least 3inches clearance. The lamp has burned out. Replace with a new Insight Instruments lamp. Refer to the lamp replacement section.
Warranty Insight Instruments warrants the Insight Instruments Solarc™ Light Source when new, to be free of defects in material and workmanship and to perform in accordance with manufacturer’s specifications when subject to normal use and service for a period of one (1) year from the date of purchase from Insight Instruments or an authorized agent.