User's Manual Part 9
Table Of Contents
Directions for Use Monitor Configuration 119
NIBP Pediatric
Systolic Upper Alarm Limit 145 mmHg 
19.3 kPa
32 - 160 mmHg
4.3 - 21.3 kPa
Monitor, Acuity
Systolic Lower Alarm Limit 75 mmHg 
10.0 kPa
30 - 158 mmHg
4.0 - 21.1 kPa
Monitor, Acuity
Diastolic Upper Alarm Limit 100 mmHg
13.3 kPa
17 - 130 mmHg
2.3 - 17.3 kPa
Monitor, Acuity
Diastolic Lower Alarm Limit 35 mmHg
4.7 kPa
15 - 128 mmHg
2.0 - 17.1 kPa
Monitor, Acuity
Mean Upper Alarm Limit 110 mmHg 
14.7 kPa
17 - 140 mmHg
2.3 - 18.7 kPa
Monitor, Acuity
Mean Lower Alarm Limit 50 mmHg 
6.7 kPa
15 - 138 mmHg
2.0 - 18.4 kPa
Monitor, Acuity
NIBP Neonatal
Systolic Upper Alarm Limit 100 mmHg 
13.3 kPa
27 - 120 mmHg
3.6 - 16.0 kPa
Monitor, Acuity
Systolic Lower Alarm Limit 50 mmHg 
6.7 kPa
25 - 118 mmHg
3.33 - 15.7 kPa
Monitor, Acuity
Diastolic Upper Alarm Limit 70 mmHg 
9.3 kPa
12 - 105 mmHg
1.6 - 14.0 kPa
Monitor, Acuity
Diastolic Lower Alarm Limit 30 mmHg 
4.0 kPa
10 - 103 mmHg
1.3 - 13.7 kPa
Monitor, Acuity
Mean Upper Alarm Limit 80 mmHg 
10.7 kPa
12 - 110 mmHg
1.6 - 14.7 kPa
Monitor, Acuity
Mean Lower Alarm Limit 35 mmHg 
4.7 kPa
10 - 108 mmHg
1.3 - 14.4 kPa
Monitor, Acuity
SpO
2
 General
Enable Spot Checks Yes Yes, No Configuration Utility
ParamSet for Upper SpO
2
5% 5% - 10% Configuration Utility
ParamSet for Lower SpO
2
5% 5% - 10% Configuration Utility
SpO
2
 Adult
Upper Alarm Limit 100% 52% - 100% Monitor, Acuity
Lower Alarm Limit 90% 50% - 98% Monitor, Acuity
SpO
2
 Pediatric
Upper Alarm Limit 100% 52% - 100% Monitor, Acuity
Lower Alarm Limit 90% 50% - 98% Monitor, Acuity
Table 11. Parameter Configuration Matrix (continued)
Parameter Factory Setting Possible Values Where Set