User's Manual Part 9
Table Of Contents
118 Monitor Configuration Welch Allyn Propaq LT Vital Signs Monitor
HR/PR Adult
Upper Alarm Limit 120 beats/min 27 - 300 beats/min Monitor
Lower Alarm Limit 50 beats/min 25 - 298 beats/min Monitor
HR/PR Pediatric
Upper Alarm Limit 150 beats/min 27 - 300 beats/min Monitor
Lower Alarm Limit 50 beats/min 25 - 298 beats/min Monitor
HR/PR Neonate
Upper Alarm Limit 150 beats/min 27 - 300 beats/min Monitor
Lower Alarm Limit 50 beats/min 25 - 298 beats/min Monitor
NIBP General
NIBP Mode Manual Auto, Manual, Turbo Monitor, Acuity
Auto Interval 15 min 1 min, 2 min, 3 min, 5 min, 10 min,
15 min, 30 min, 60 min
Monitor, Configuration
Utility
ParamSet for Upper NIBP
Systolic
15% 5% - 25% Configuration Utility
ParamSet for Lower NIBP
Systolic
15% 5% - 25% Configuration Utility
ParamSet for Upper NIBP
Diastolic
15% 5% - 25% Configuration Utility
ParamSet for Lower NIBP
Diastolic
15% 5% - 25% Configuration Utility
ParamSet for Upper NIBP
Mean
10% 5% - 15% Configuration Utility
ParamSet for Lower NIBP
Mean
10% 5% - 15% Configuration Utility
NIBP Adult
Systolic Upper Alarm Limit 220 mmHg
29.3 kPa
32 - 260 mmHg
4.3 - 34.7 kPa
Monitor, Acuity
Systolic Lower Alarm Limit 75 mmHg
10.0 kPa
30 - 258 mmHg
4.0 - 34.4 kPa
Monitor, Acuity
Diastolic Upper Alarm Limit 110 mmHg
14.7 kPa
22 - 235 mmHg
2.9 - 31.3 kPa
Monitor, Acuity
Diastolic Lower Alarm Limit 35 mmHg
4.7 kPa
20 - 233 mmHg
2.7 - 31.1 kPa
Monitor, Acuity
Mean Upper Alarm Limit 120 mmHg
16.0 kPa
22 - 255 mmHg
2.9 - 34.0 kPa
Monitor, Acuity
Mean Lower Alarm Limit 50 mmHg
6.7 kPa
20 - 253 mmHg
2.7 - 33.7 kPa
Monitor, Acuity
Table 11. Parameter Configuration Matrix (continued)
Parameter Factory Setting Possible Values Where Set