User's Manual Part 9

Directions for Use Monitor Configuration 117
Display 2
Format Single Waveform Large Numerics, Single Waveform,
Dual Waveform, Tabular Trends,
Tabular Trends with Waveform
Monitor, Configuration
Utility
Top Waveform (if any) Lead II Lead I, Lead II, Lead III, Lead V, aVR, aVL,
aVF, SpO
2
, Resp
Monitor, Configuration
Utility
Show Top Waveform 6
Seconds
No Yes, No Monitor, Configuration
Utility
Show Bottom Waveform 6
Seconds
No Yes, No Monitor
Bottom Waveform (if any) Lead V Lead I, Lead II, Lead III, Lead V, aVR, aVL,
aVF, SpO
2
, Resp
Monitor, Configuration
Utility
Tabular Trends Display
Interval
5 min 1 min, 5 min, 10 min, 15 min, 30 min,
60 min
Monitor, Configuration
Utility
Display 3
Format Tabular Trends with
Waveform
Large Numerics, Single Waveform,
Dual Waveform, Tabular Trends,
Tabular Trends with Waveform
Monitor, Configuration
Utility
Top Waveform (if any) Lead II Lead I, Lead II, Lead III, Lead V, aVR, aVL,
aVF, SpO
2
, Resp
Monitor, Configuration
Utility
Show Top Waveform 6
Seconds
No Yes, No Monitor, Configuration
Utility
Show Bottom Waveform 6
Seconds
No Yes, No Monitor
Bottom Waveform (if any) SpO
2
Lead I, Lead II, Lead III, Lead V, aVR, aVL,
aVF, SpO
2
, Resp
Monitor, Configuration
Utility
Tabular Trends Display
Interval
5 min 1 min, 5 min, 10 min, 15 min, 30 min,
60 min
Monitor, Configuration
Utility
ECG
ECG Bandwidth Monitor Monitor, Extended Monitor, Configuration
Utility, Acuity
HR/PR General
HR/PR Selected Source ECG ECG, SpO
2
Monitor, Acuity
Can Disable HR/PR Alarm
Limits
Yes
(No if lang=French)
Yes, No Monitor
Power Source Filter 60 Hz 50 Hz, 60 Hz, Off Monitor, Configuration
Utility, Acuity
ParamSet for Upper HR/PR
Limit
20% 5% - 25% Configuration Utility
ParamSet for Lower HR/PR
Limit
20% 5% - 25% Configuration Utility
Table 11. Parameter Configuration Matrix (continued)
Parameter Factory Setting Possible Values Where Set