User's Manual Part 9

Configuration Worksheet 151
7 Monitor Behavior
For each parameter, the factory value is shown in bold. The factory value is used until
another value is uploaded to the monitor.
7.1 Patient Mode Default
In what default patient mode should the monitor be set for new patients?
__ Adult
__ Pediatric
__ Neonatal
7.2 Display All ECG Vectors
If a 5-lead ECG cable is connected to the monitor, should the monitor be able to
display the optional aV
R
, aV
L
, and aV
F
waveforms?
(Yes)
(No)
7.3 ECG Bandwidth Default
ECG Bandwidth refers to the frequency range used to display and print patient
data.
Monitor bandwidth is less susceptible to artifact such as patient motion.
Extended bandwidth, although more susceptible to artifact, can provide for more
detailed data analysis.
Which ECG bandwidth should be displayed by default?
__ Monitor
__ Extended
Monitor-Mode Bandwidth
Adult 0.5 Hz to 40 Hz
Pediatric/Neonate 0.5 Hz to 80 Hz
Extended-Mode Bandwidth
Adult 0.05 Hz to 40 Hz
Pediatric/Neonate 0.05 Hz to 80 Hz