User's Manual Part 9

114 Monitor Configuration Welch Allyn Propaq LT Vital Signs Monitor
Table 11. Parameter Configuration Matrix
Parameter Factory Setting Possible Values Where Set
General
Patient Name
(Last, First, MI)
[blank] (16 characters displayed; actual
number of characters depends on
character width)
Monitor, Acuity
Patient ID [blank] (12 characters) Monitor, Acuity
Room [blank] (5 characters) Monitor, Acuity
Regulatory Setting US US, European EC, Japan Metrology Configuration Utility,
Acuity
Patient Mode Adult Adult, Pediatric, Neonate Monitor, Configuration
Utility, Acuity
LCD Bias 32 1 - 64 Monitor
Network
Radio Card ON and Acuity
Enabled
YES if the radio
card is installed.
NO if the radio card
is not installed.
Yes, No Configuration Utility
Radio ESSID (NetName) com.protocol com.protocol(1), demo.protocol(2)
com1.protocol(3), com2.protocol(4)
com1.protocol(5), com2.protocol(6)
com1.protocol(7), com2.protocol(8)
com1.protocol(9), com2.protocol(10)
com1.protocol(11), com2.protocol(12)
com1.protocol(13), com2.protocol(14)
com1.protocol(15), com2.protocol(16)
com1.protocol(17), com2.protocol(18)
com1.protocol(19), com2.protocol(20)
Monitor
CAUTION Do not attempt to change the network name unless you are a qualified
biomedical service engineer or technician. Changing the network name causes
the monitor to disconnect from the current FlexNet network, restart (deleting all stored
patient data), and attempt to connect to the FlexNet network corresponding to the new
network name.
To change the network name:
1. Enter the Service menu.
2. Highlight Radio.
3. Highlight NetName and press .
4. Select another network name and press .
5. Press to return to the monitoring display.
Radio Disconnect Not Disconnected Disconnected, Not Disconnected Monitor