User's Manual Part 9

148 Welch Allyn Propaq LT Vital Signs Monitor
6.2 Enable Lockouts
If lockouts are enabled, simultaneously pressing the Left Arrow, Right Arrow, and
Down Arrow buttons locks out one or more of the following:
For the following questions, select which lockouts, if any, to enable for this monitor
configuration.
6.2.1 Buttons Lockout Enable
Buttons lockout prevents unauthorized personnel from operating the monitor.
If you enable buttons lockout, the 3-button combination (Up Arrow, Down Arrow,
and Right Arrow) locks out the monitor until the combination is repeated or until an
alarm or alert condition arises.
Do you want to enable buttons lockout on this monitor?
(Yes)
(No)
6.2.2 Display Lockout Enable
Display lockout enable prevents unauthorized viewing of patient vital signs.
If display lockout is enabled, the 3-button combination (Up Arrow, Down Arrow, and
Right Arrow) locks out the monitor display until the combination is repeated or until
an alarm or alert condition arises.
Do you want to enable display lockout on this monitor?
(Yes)
(No)
6.2.3 Back Light Lockout Enable
Back light lockout prevents unauthorized viewing of patient vital signs.
If back light lockout is enabled, the 3-button combination (Up Arrow, Down Arrow,
and Right Arrow) locks out the monitor display back light until the combination is
repeated or until an alarm or alert condition arises.
Do you want to enable back light lockout on this monitor?
(Yes)
(No)
Buttons Block access to the monitor buttons
Back Light Extend the battery charge life
Display Prevent unauthorized viewing of confidential patient information