User's Manual Part 9
Table Of Contents
136 PC Utility Welch Allyn Propaq LT Vital Signs Monitor
After you reconfigure the monitor, the settings defined in the new configuration become 
the ‘system defaults’, and the factory defaults are no longer in effect. The new system 
defaults remain in effect until another configuration is downloaded to the monitor. If you 
reconfigure the monitor and then want to restore the factory defaults, download the 
configuration file “DefaultPropaqLTConfiguration.mnt.”
The monitor is reset to the system defaults (not the factory defaults) every time the 
monitor power is cycled off and on, with one exception
: If a clinician has changed some 
monitor settings at the bedside for a patient and then shuts off the monitor, the monitor 
offers the options to “Delete and Shut Down” and to “Save and Shut Down”. If the 
clinician selects “Save and Shut Down,” shuts off the monitor, and then turns on the 
monitor again, the monitor prompts with the choice, “Start New Patient” or “Continue 
Patient”. If the clinician selects “Continue Patient,” the monitor uses the settings that 
were saved for that patient before the power was turned off. If the clinician selects “Start 
New Patient,” the system default settings are restored.
• In the printed version of the configuration worksheet, the factory default values 
are indicated by bold text. 
• In the electronic version of the configuration worksheet, the factory default 
values appear in the form fields when you first open the file, 
PropaqLTMonCfgWS.doc.
• In the configuration utility, the factory default values appear in the data fields 
when you open the default configuration file, DefaultPropaqLTConfiguration.mnt, 
using the configuration utility.
About AutoPrint
For information about using the AutoPrint Utility, see “Printing Patient Data” on page 103.
Caution Do not attempt to modify a configuration file without using the Propaq 
LT Configuration Utility.