User's Manual Part 8

104 Printing Patient Data Welch Allyn Propaq LT Vital Signs Monitor
The printed data includes the following:
Trends print-out:
Up to 24 hours of HR/PR, RR/BR, and SpO
2
trends (one set of measurements per
row, with a 15-minute interval between rows).
•SpO
2
spot checks and NIBP measurements, interspersed among the rows of
trends data.
In the Snapshots print-out:
Up to twenty snapshots (21-second intervals of full-disclosure patient data,
including waveforms and numerics).
The utility extracts the available ECG Leads from the patient data and prints a
page of strips for each snapshot. Depending on the number of ECG Leads
available, the strips may contain 21 seconds of data for one or three Leads of
ECG data.
The utility supports US Letter and European A4 page sizes.
Printing
To print the patient data stored in a monitor:
1. Install the AutoPrint Utility on a PC.
a. Insert the Propaq LT Monitor PC Utility Program in the CD-ROM drive of the PC.
b. Follow the on-screen instructions to install the AutoPrint Utility.
2. Connect the PC to a cradle (Propaq LT Monitor Charging/Communications Cradle),
using the supplied USB cable.
Each time a monitor is inserted into this cradle while the cradle is connected via USB
cable to the PC, the patient data stored in the monitor is printed on the default printer.
Patients monitored continuously for 24 hours typically generate enough data for a 2-
or 3-page trends table of two panels per page. For example, a 24-hour report with the
following data (192 rows) would cover three pages:
96 rows of vital-signs data (4 intervals/hr = 4 rows per hour) for 24 hours
48 NIBP readings (2/hr)
•48 SpO
2
spot-check readings (2/hr)