User's Manual Part 7
Table Of Contents
Directions for Use Alarms and Alerts 93
NIBP Fault.
No valid blood pressure found.
The patient mode setting is incorrect or the wrong hose or 
cuff is being used for the current patient mode.
NIBP Fault.
Calibrating. 
Please wait.
The monitor periodically calibrates (zeroes) the NIBP 
channel to make sure it can properly make NIBP 
measurements. No NIBP monitoring can be done until the 
calibration is completed. Other normal monitor operation 
continues during NIBP calibration.
NIBP Fault.
Calibrating. 
Minimize motion.
Motion is detected during a periodic NIBP calibration. 
Minimize patient motion or motion on the cuff, or 
disconnect the cuff. Motion-generated noise on the 
pressure transducer can cause the calibration to continue 
indefinitely.
NIBP Fault.
Low battery. 
NIBP disabled.
The battery is too far discharged to operate the NIBP 
channel. Insert the monitor into a powered cradle.
NIBP Fault.
Service required. 
NIBP disabled.
Have the monitor serviced.
NIBP Fault.
Kinked or neonate hose.
A hose is kinked or a neonate hose is detected in the adult 
patient mode. Check the hose and the patient mode 
selection.
NIBP Fault.
Artifact present.
Minimize motion.
The monitor has detected too much artifact to allow 
accurate readings. Take steps to reduce artifact. Position 
the patient’s limb away from the body so the applied cuff is 
not in contact with the patient’s body or any other object 
such as a bed rail.
Network 
Communication 
Comms Fault.
Check Acuity/network connection.
The monitor detects a network communication problem.
Comms Fault.
Check USB connection.
The monitor detects a problem in communication with the 
cradle. Detach and reattach the USB cable.
Not on Network.
Patient info entry not allowed.
You attempted to select the Name, ID, or Rm field on an 
Acuity-enabled monitor that is not connected to the 
network.
Battery Low Battery.
Charge battery soon.
The monitor battery charge is low, and the monitor will shut 
down in 30 minutes or less. Insert the monitor into the 
cradle. If no cradle is available, find an alternative method 
of monitoring the patient before the monitor shuts down.
Very Low Battery.
Charge battery now.
The monitor battery charge is very low; the monitor will shut 
down in 5 minutes or less. Insert the monitor into a cradle or 
find another way to monitor the patient before the monitor 
shuts down.
Battery Too Low.
Shutting down.
The monitor battery charge is too low to support monitor 
function. Monitor operation can continue only after the 
battery is recharged or replaced or until the monitor is 
inserted in a powered cradle.
Table 9. Alert Messages and Status Messages (continued)
Alert Type Message Possible Cause and Suggested Response










