User's Manual Part 7
Table Of Contents
92 Alarms and Alerts Welch Allyn Propaq LT Vital Signs Monitor
Alert Messages and Status Messages
Table 9. Alert Messages and Status Messages
Alert Type Message Possible Cause and Suggested Response
ECG ECG Fault.
XX lead failed.
Lead XX (LA, LL, RA, C, or RL) has very poor contact or no 
contact with the patient. Check for proper connection and 
replace the electrode if needed.
ECG Fault.
XX, XX leads failed.
Leads XX and XX (any two leads on a 5-lead cable) have very 
poor contact or no contact with the patient. Check for proper 
connection; replace electrodes if needed.
ECG Fault.
Multiple lead fail.
At least three leads of a 5-lead cable or at least two leads 
of a 3-lead cable have very poor contact or no contact with 
the patient. Check for proper connection; replace electrodes 
if needed.
ECG Fault.
Excessive offset.
At least one channel has excessive offset. At least one 
electrode is old, contaminated, or defective. Replace the 
electrodes.
ECG Fault.
Cable disconnected.
The ECG cable is unplugged.
NIBP NIBP Fault.
Air leak. 
Check hose.
The monitor could not properly inflate the cuff. Check the 
hose and cuff for leaks.
NIBP Fault.
Kinked hose.
Check hose.
The monitor could not properly inflate the cuff. Check for a 
hose kink between the monitor and the patient.
NIBP Fault.
Overpressure condition.
The pressure in the cuff exceeded the acceptable limits for 
the current patient mode. Check the hose and retry the 
measurement.
NIBP Fault.
Weak Pulses. 
Can’t find Sys/Dia.
Not enough pulses to determine the systolic or diastolic 
pressures, but a mean pressure is available. Squeeze all air 
from the cuff and reapply it.
NIBP Fault.
Artifact. 
Can’t find Sys/Dia.
The systolic or diastolic pressures are unreliable due to 
artifact, but a mean pressure is available. Usually caused by 
patient motion.
NIBP Fault.
No pulses detected.
The cuff might not be properly applied to the patient, or the 
patient might not have detectable pulses due to shock or 
arrhythmias.
WARNING The monitor cannot determine whether 
this alert has a physiologic cause or a cuff 
application cause. Always evaluate the patient for 
presence of life
-threatening conditions when this 
message occurs.
NIBP Fault.
Connect ECG to reduce NIBP artifact.
NIBP artifact prevents a valid reading. Connect ECG 
electrodes to improve NIBP measurements. (See “Improving 
NIBP Accuracy with Smartcuf” on page 71.)










