User's Manual Part 7
Table Of Contents
Directions for Use Alarms and Alerts 91
To Respond at the Monitor to an Alert
1. Press  to silence the tone for 90 seconds; press  to clear the alert.
2. Troubleshoot and correct the problem. (See “Alert Messages and Status Messages” 
on page 92.)
To Respond at Acuity to an Alert
For a monitor in communication with Acuity, alarms and alerts can be detected by either 
the monitor or Acuity, and are displayed in both places. See Acuity Directions For Use.
About Battery Charge Status
Table 8. Battery Status Indicators
Display Status/Alert
Text
Battery Monitor Recommended 
Action
Green
(none) Fully charged Okay None
Green
Partially full Okay None
Green
Partially full; charging Okay None
Yellow
EQUIPMENT ALERT
Low Battery.
Charge battery soon.
Low Can function for up to 
30 minutes, but NIBP 
is disabled.
Prepare to discontinue 
monitoring. If possible, 
insert the monitor into a 
cradle.
Yellow
Low; charging Normal function. Do not remove the 
monitor from the cradle.
Red
EQUIPMENT ALERT
Battery Too Low.
Shutting down.
Almost completely 
discharged
Shutting down soon. 
Can function for up to 
5 minutes, but NIBP is 
disabled. 
Prepare to discontinue 
monitoring. If possible, 
insert the monitor into a 
cradle.
If this indicator appears 
when the monitor is in a 
powered cradle, then 
the battery is damaged 
and must be replaced.
In this case, all stored 
patient data will be 
deleted when the 
monitor is removed 
from the cradle.
Red
Almost completely 
discharged; charging
Normal function.  Do not remove the 
monitor from the cradle.