User's Manual Part 7

88 Alarms and Alerts Welch Allyn Propaq LT Vital Signs Monitor
Figure 86. Suspend Audible Alarms: Off
3. Highlight On (Figure 87).
Figure 87. Suspend Audible Alarms: On
The alarm tone is suspended immediately.
If an alarm condition occurs during the suspension period, the alarm tone does
not sound.
A count-down timer appears below the line to indicate the time remaining
in the
suspension period.
Audio alarms suspended’ appears in yellow in the upper left corner of the screen.
When the suspension period elapses, the alarm tone is again enabled.
Note
If audible alarms are suspended, pressing cancels the suspension.
II 1mV/cm
II 1mV/cm
Exit
Exit
Trends
Trends
Sna
p
shots
Snapshots
Setu
pSetup
Low
Low
Med
Med
HR/min
HR/min
NIBP mmHg
NIBP mmHg
Resp/min
Resp/min
SpO2
SpO2
Off
Off
On
On
Alarm Tone
Alarm Tone
Suspend Audible Alarms
Suspend Audible Alarms
Setup
Setup
(90 sec)
(90 sec)
%
3:00:06P
3:00:06P
Adult
Adult
Rm 239
Rm 239
Alarms
Alarms
ECG
ECG
NIBP
NIBP
Timings
Timings
High
High
Service
Service
HALL, ROBERT E.
HALL, ROBERT E.
3456187
3456187
Right ( ) to suspend audible alarms.
Right ( ) to suspend audible alarms.
II 1mV/cm
II 1mV/cm
Exit
Exit
Trends
Trends
Snapshots
Snapshots
Setup
Setup
Low
Low
Med
Med
80
80
HR/min
HR/min
NIBP mmHg
NIBP mmHg
Resp/min
Resp/min
SpO2
SpO2
Off
Off
On
On
Alarm Tone
Alarm Tone
Suspend Audible Alarms
Suspend Audible Alarms
Setup:
Setup:
(90 sec)
(90 sec)
12
12
97
97
%
8765432
8765432
3:00:06P
3:00:06P
Adult
Adult
Rm 239
Rm 239
Alarms
Alarms
ECG
ECG
NIBP
NIBP
Timings
Timings
High
High
Service
Service
0:01:30 remaining
0:01:30 remaining
Audio alarms suspended
Audio alarms suspended
ID: 01018VDO9PBH
ID: 01018VDO9PBH
‘Tone suspended’ indicator
Suspension time remaining
Configured suspension period