User's Manual Part 4

36 Overview of Monitor Operation Welch Allyn Propaq LT Vital Signs Monitor
About the NIBP Control Menu
The NIBP control (Figure 28) has four submenus: Manometer, Systolic, Diastolic, and
Mean. Press or to select a submenu.
Figure 28. NIBP Control Menu
Note
For manometer information, see “To Use the Digital Manometer” on page 73.
Parameter Options Parameter Options
Systolic Systolic
Upper Sys Alarm Off On Lower Sys Alarm Off On
Upper Sys Limit Lower Sys Limit
Adult
Pediatric
Neonate
32 - 260 mmHg
32 - 160 mmHg
27 - 120 mmHg
Adult
Pediatric
Neonate
30 - 258 mmHg
30 - 158 mmHg
25 - 118 mmHg
NIBP Mode Auto Manual Turbo Auto Interval (min) 123510153060
Diastolic Diastolic
Upper Dia Alarm Off On Lower Dia Alarm Off On
Upper Dia Limit Lower Dia Limit
Adult
Pediatric
Neonate
22 - 235 mmHg
17 - 130 mmHg
12 - 105 mmHg
Adult
Pediatric
Neonate
20 - 233 mmHg
15 - 128 mmHg
10 - 103 mmHg
NIBP Mode Auto Manual Turbo Auto Interval (min) 123510153060
MAP MAP
Upper MAP Alarm Off On Lower MAP Alarm Off On
Upper MAP Limit Lower MAP Limit
Adult
Pediatric
Neonate
22 - 255 mmHg
17 - 140 mmHg
12 - 110 mmHg
Adult
Pediatric
Neonate
20 - 253 mmHg
15 - 138 mmHg
10 - 108 mmHg
NIBP Mode Auto Manual Turbo Auto Interval (min) 123510153060
II 1mV/cm
II 1mV/cm
1
1
2
2
3
3
5
5
10
10
15
15
30
30
60
60
Exit
Exit
Trends
Trends
Snapshots
Snapshots
Setup
Setup
Auto
Auto
Manual
Manual
Turbo
Turbo
HR/min
HR/min
NIBP mmHg
NIBP mmHg
Resp/min
Resp/min
SpO2
SpO2
Off
Off
Off
Off
On
On
On
On
220
220
Lower Sys Limit
Lower Sys Limit
NIBP Mode
NIBP Mode
Auto Interval (min)
Auto Interval (min)
Upper Sys Alarm
Upper Sys Alarm
Upper Sys Limit
Upper Sys Limit
Lower Sys Alarm
Lower Sys Alarm
NIBP Manometer
NIBP Manometer
Systolic
Systolic
Diastolic
Diastolic
Mean
Mean
75
75
%
%
STEWART, ANN
STEWART, ANN
7762940
7762940
3:00:06P
3:00:06P
Adult
Adult
Rm 263
Rm 263
(102)
(102)
80
80
140/78
140/78
12
12
97
97