User's Manual Part 3

Directions for Use Overview of Monitor Operation 23
Timing Out the Display and the Back Light
To conserve battery run life, the monitor can be configured to turn off the display and the
back light if no operator activity (that is, a button press) is detected for a specified number
of minutes.
When the display is off, no patient data is visible.
When the back light is off, patient data is visible only under direct light.
The monitor configuration determines whether the time-out feature is enabled or
disabled, and defines the default time-out period if it is enabled. If the feature is enabled in
the configuration, you can temporarily change the time-out period or disable the time out
through the Setup -> Timings menu.
Locking Out the Display, Back Light, and Buttons
If lock-outs are enabled, you can lock out one or more of the following elements:
Buttons To prevent unauthorized use
Back Light To extend the battery run life
Display To prevent unauthorized viewing of patient information
The monitor configuration defines which of these, if any, can be locked out. For
information on enabling or disabling lock-outs for these elements, see “Monitor
Configurationon page 111.
To Lock Out the Configured Elements
Hold down , , and simultaneously for 5 seconds.
To Unlock the Configured Elements
Hold down , , and simultaneously for 5 seconds.
Note
If the display and the back light are turned off due to a time out, they turn on again
immediately when an alarm or alert occurs or, if the buttons lock-out is not
enabled, when you press any monitor button.
Note
Alarms and alerts immediately unlock any locked elements.