User's Manual Part 2
Table Of Contents
Directions for Use Introduction 5
Safety
The monitor is safe for patients and clinicians when used in accordance with the 
instructions and with the warning and caution statements presented in this manual.
All personnel must read and understand all warning and caution statements 
presented in this manual before using the monitor. 
• Failure to understand and observe any warning statement in this manual could 
lead to patient injury, illness, or death.
• Failure to understand and observe any caution statement in this manual could 
lead to equipment damage or loss of patient data.
General Warnings
These statements apply to all aspects of patient monitoring. Statements which apply 
specifically to one aspect of monitoring, such as NIBP or SpO
2
 monitoring, are presented 
in the corresponding sections of the manual.
WARNING Many environmental variables, including patient physiology and 
clinical application, can affect the accuracy and performance of the monitor. The 
clinician must verify all vital-signs information prior to patient intervention.
WARNING Always check the patient mode (adult, pediatric, or neonate) when 
monitoring a new patient. The patient mode determines default alarm limits and 
internal algorithm settings. Make sure the monitor has settings that are 
appropriate before monitoring the patient.
WARNING Make sure Acuity patients, and especially those prone to 
arrhythmias, are kept under close surveillance. While monitoring patients with 
Acuity, the clinician must review all clinical data before implementing therapy. As 
with all computerized arrhythmia analysis systems, Acuity cannot replace skilled 
care and proper surveillance by a clinician.
WARNING It is possible for Acuity alarms, alerts, or other events to go 
unnoticed if clinical personnel are not present at Acuity or if interruptions occur in 
power or system operations. To help reduce this possible occurrence, Acuity 
must be installed with redundant power supplies and redundant means of 
operator surveillance, such as secondary Acuity Central Stations and hallway 
message panels.
WARNING The monitor might not meet its performance specifications if stored 
or used outside the specified temperature and humidity ranges.
WARNING Use of respiration monitoring by impedance pneumography can 
affect the operation of some pacemakers. If pacemaker operation is affected, turn 
off respiration pneumography. (See Figure 61 on page 61.)
WARNING Do not connect more than one patient to a monitor. 
WARNING Do not connect more than one monitor to a patient.