162 Maintenance Welch Allyn Propaq LT Vital Signs Monitor
Recycling Monitor Components
Within the EU
Do not dispose of this product as ‘unsorted municipal waste’. Prepare it for reuse 
or separate collection as specified by Directive 2002/96/EC of the European 
Parliament and the Council of the European Union on Waste Electronic and 
Electrical Equipment (WEEE).
For more specific disposal information, see www.welchallyn.com/weee, or contact Welch 
Allyn Customer Service.
Outside the EU
When the monitor, the cradle, or the battery reaches end of life, recycle it locally 
according to national, state, and local regulations, or return it to Welch Allyn.
Caution Never use any of the following solutions or similar products to clean 
the equipment: butyl alcohol, denatured ethanol, Freon™, mild chlorine bleach 
solution, isopropyl alcohol (except for the SpO
2
 cable), trichloroethane, 
trichloroethylene, acetone, Vesphene II, Enviroquat
®
, Staphene
®
, Misty
®
, 
Glutaraldehyde, Formula 409
®
, or Fantastik
®
.
Table 12. Cleaning Instructions and Cleaning Solutions
Equipment Cleaning Instructions Approved Cleaning Solutions
Monitor
a
Charging/
Communication Cradle
a
a. The equipment can be disinfected to comply with OSHA requirements for cleaning and decontaminating spills of blood and other body 
fluids. (Federal OSHA Standard on bloodborne pathogens: 29 CFR 1910.1030, 12/6/91.)
• Wipe with a nearly dry cloth moistened with 
cleaning solution.
• Thoroughly wipe off any excess cleaning 
solution. Do not let cleaning solution run into 
connector openings or crevices.
b
b. If liquid gets into the connectors, dry the area with warm air and then verify all monitoring functions.
Warm water, liquid soap, Coverage
®
, 
Windex
®
, Ovation
®
, hydrogen 
peroxide solution, Wex
-cide
®
c
, 
T.B. Q.
®
c
c. Wex-cide (Wexford Labs, Inc., Kirkwood, MO) and T.B.Q. (Calgon Vestal Lab., Calgon Corp., St. Louis, MO) are disinfectants that meet 
OSHA requirements, are EPA approved, and will not harm the outside of the monitor, battery, or battery charger. Wipe away 
disinfectants with a water
-dampened cloth after the manufacturer’s recommended period.
ECG cable, extension 
cable
• Wipe gently with damp cloth moistened with a 
mild detergent solution.
• Thoroughly wipe off any cleaning solution.
Mild detergent.
SpO
2
 cable, extension 
cable
• Wipe the cable with a 70% isopropyl alcohol pad 
and allow it to dry.
70% isopropyl alcohol pad.
Other accessories • Consult manufacturer’s instructions. Consult manufacturer’s instructions.
Note
If the monitor, the cradle, or the battery is contaminated, this directive does not 
apply.