User's Manual Part 10
Table Of Contents
184 EMC Compliance Welch Allyn Propaq LT Vital Signs Monitor
Charging/Communication Cradle with Monitor
Guidance and manufacturer’s declaration—electromagnetic emissions
The Charging/Communication Cradle with Propaq LT Series monitor is intended for use in the electromagnetic
environment specified below. The customer or the user of the Charging/Communication Cradle with Propaq LT Series
monitor should assure that it is used in such an environment.
Note: The EMC specifications on pages 184-187 apply to the following:
• Charging/Communication Cradle alone
• Propaq LT Series monitor with Charging/Communication Cradle
• Propaq LT Series monitor with Charging/Communication Cradle in communication via USB cable with personal
computer
Emissions test Compliance Electromagnetic environment—guidance
RF emissions
CISPR 11
Group 1 The Charging/Communication Cradle with Propaq LT Series monitor
uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions
CISPR 11
Class B The Charging/Communication Cradle with Propaq LT Series monitor
is suitable for use in all establishments, including domestic
establishments and those directly connected to the public
low
-voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000
-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000
-3-3
Complies