User Manual Part 5

52 Standalone Monitoring Welch Allyn Propaq LT Vital Signs Monitor
6. Highlight Confirm and press .
\
7. Verify that the patient mode (adult, pediatric, or neonate) is set correctly for this
patient.
Neonatal Term birth through 28 days, or up to 44 gestational weeks.
Pediatric Between 29 days and 12 years.
Adult 13 years and older.
If the current patient mode setting is not correct:
a. Highlight the current patient mode (Adult, Pediatric, or Neonate) and press .
b. From the Patient Mode selection menu (Figure 48), highlight the appropriate
patient mode and press .
Figure 48. Changing Patient Mode
c. A confirmation screen appears:
Figure 49. Confirming a Change of Patient Mode
d. To confirm the change, highlight Confirm and press .
Note
Name alone is not sufficient to confirm a patient ID; thus, you can confirm the
patient name only after you have entered the patient ID.
Note
When you change the patient mode and confirm the change:
All vital-signs data for the patient is lost.
All monitor settings revert to the defaults for the new patient mode.
Patient Mode
Adult
Pediatric
Neonate
Patient Mode
Patient Mode
Adult
Adult
Pediatric
Pediatric
Neonate
Neonate
HALL, ROBERT E.
HALL, ROBERT E.
3456187
3456187
3:00:06P
3:00:06P
Adult
Adult
Rm 239
Rm 239
NIBP mmHg
NIBP mmHg
Resp/min
Resp/min
HR/min
HR/min
SpO2
SpO2
%
S D M
Note! You are changing the
Note! You are changing the
patient mode from Adult to Pediatric.
patient mode from Adult to Pediatric.
This will change settings to default
This will change settings to default
Pediatric values and delete previous
Pediatric values and delete previous
If this is correct, highlight “Confirm”
If this is correct, highlight Confirm
then press Select ( ).
then press Select ( ).
Cancel
Cancel
Confirm
Confirm
patient data.
patient data.