User Manual Part 5

74 Standalone Monitoring Welch Allyn Propaq LT Vital Signs Monitor
Figure 73. NIBP: Manometer Reading Complete
NIBP Measurements in Power-Saving Mode
When a manual or turbo NIBP activity awakens the monitor from power-saving mode
(“Power Saving” on page 45), cuff inflation pressure is reset to default levels and cuff
inflation is delayed for up to 8 seconds.
NIBP Disabled When the Battery is Low
If the battery is low and the monitor is operating on battery power, NIBP functions are
disabled and the monitor displays the message “NIBP off. Low battery.
If you attempt to start an NIBP measurement during a low-battery condition, the monitor
displays an equipment alert with the message “Low battery. NIBP disabled.
NIBP Mode:
NIBP Mode:
II 1mV/cm
II 1mV/cm
Exit
Exit
Trends
Trends
Snapshots
Snapshots
Setup
Setup
PR/min
PR/min
NIBP mmHg
NIBP mmHg
Resp/min
Resp/min
SpO2
SpO2
NIBP
NIBP
Systolic
Systolic
Diastolic
Diastolic
Mean
Mean
%
%
STEWART, ANN
STEWART, ANN
7762940
7762940
3:00:06P
3:00:06P
Adult
Adult
Rm 263
Rm 263
Manometer
Manometer
Manual
Manual
2
2
50
50
100
100
150
150
200
200
250
250
300
300
(98)
(98)
79
79
132/83
132/83
Systolic, diastolic, and MAP
(numeric)
Systolic, diastolic, and MAP
(graphical)
Note
Inserting a monitor into a powered cradle during a low-battery condition
immediately enables NIBP monitoring.