Improving the Waveform Display
If the power source filter is off, noise from the power source can cause an unclear or
noisy waveform.
To Change the Power Source Filter to Reduce Noise
1. Access the Setup menu. (See “To Access the Setup Menus” on page 38.)
2. Highlight ECG.
Verify that the setting for Power source filter (Figure 62) is correct for the power
source in your facility. If you do not know what this setting should be, consult a
qualified service person.
3. Highlight Power source filter and press or as needed to select 60 Hz, 50 Hz,
or Off.
Figure 62. Turning On the 60 Hz Power Source Filter in the ECG Setup Menu
4. Exit the Setup menu by pressing or .
Monitoring Respiration
Resp is based on impedance pneumography, where respirations are sensed from the
ECG electrodes.
Resp is part of ECG monitoring. The Resp numeric is displayed (in purple) in the lower
right corner. To view the Resp waveform:
Change the waveform source to Resp (Figure 63). (See “To Change the Waveform
Selectionon page 59.)
II 1mV/cm
II 1mV/cm
Exit
Exit
Trends
Trends
Snapshots
Snapshots
Setup
Setup
60Hz
60Hz
50Hz
50Hz
80
80
HR/min
HR/min
140/78
140/78
NIBP mmHg (102)
NIBP mmHg (102)
Resp/min
Resp/min
SpO2
SpO2
Off
Off
On
On
Power source filter
Power source filter
Resp Monitoring
Resp Monitoring
Setup
Setup
12
12
97
97
%
HALL, ROBERT E.
HALL, ROBERT E.
3456187
3456187
3:00:06P
3:00:06P
Adult
Adult
Rm 239
Rm 239
Alarms
Alarms
ECG
ECG
NIBP
NIBP
Timings
Timings
Off
Off
Service
Service
ECG Bandwidth
ECG Bandwidth
Pacer indicator
Pacer indicator
Monitor
Monitor
Off
Off
On
On
Extended
Extended
Power source filter 60 Hz
Note
All ECG cables listed for the Propaq LT monitor in Products and Accessories
(810-0409-XX) permit respiration monitoring and electrosurgical interference
suppression.
Note
To measure Resp with Lead II selected, the LL lead must be attached to the
patient.
WARNING Use of respiration monitoring by impedance pneumography can
affect the operation of some pacemakers. If pacemaker operation is affected, turn
off respiration pneumography. (See Figure 61 on page 61.)