Product Manual

97
APPENDIX A
Custom ID Download
Custom ID formats are uniquely defined by your facility’s needs. This customized ECG header information is
designed in ELI Link and downloaded to the device.
From real-time ECG view, select F6 (More) followed by F2 (Custom ID Download). “Transmission Status”
will remain visible for approximately 10 seconds followed by “Waiting for Response”, “Connected”, and
“Custom ID downloaded”. A return to the real-time ECG view indicates the custom ID download is complete.
The custom ID remains the new header format for all future ECGs until you select a different ID format in the
configuration settings. You may alter the ID format configuration to short, standard, long, or custom based on
your patient demographic entry needs. The custom ID is only deleted upon downloading a new custom ID or
on the rare occasion of downloading software – it will not be lost due to power loss or switching to a different
ID format.
TIP: Upon custom ID download, the ID format will assume the group name as designed in ELI Link
or E-Scribe.
NOTE: The site number must be configured in the electrocardiograph and recognized as an
established, valid site number at the E-Scribe before downloading the custom ID.
TIP: Confirm the baud rate in the configuration settings before downloading the custom ID from ELI
Link or E-Scribe (applies to ELI 150c only).
USB Memory
Both the ELI 150c and ELI 250c come equipped with a standard USB host port which can be used to transmit
patient records from the device’s internal memory to an external USB memory stick. In addition, either device
can be equipped with an optional USBD (device) port. The optional USBD port can be used to directly connect
the device to a PC running ELI Link V3.10 or later.
Transmission using the USB host port to a USB memory stick
USB memory communication media permits storage of patient records to an external USB memory stick. The
files will be saved in UNIPRO32 format for transfer to the E-Scribe or a compatible electronic information
management system.
NOTE: Device is compatible with FAT32 formatted USB memory sticks.
NOTE: The USB memory stick must not contain any automatic features (e.g., Sandisk U3). Uninstall any
features from the memory stick before connecting it to the device.
NOTE: All communication options (MODEM, LAN, WLAN, GPRS) including time sync and orders
download are disabled when a USB memory stick is inserted in the device.
NOTE: Upon successful transmission the device display will state “Transmission status transmit
complete. Press any key to continue.” The total number of patient records transmitted to the USB
memory stick will also appear.
NOTE: Patient records transmitted to a USB memory stick are marked as transmitted by the device.
Transferring Individual Patient Records to the USB Memory Stick
Place the USB memory stick into the USB host port on the back of the device.
Select F6 (More) from real-time ECG view.
Select F1 (Directory of Stored ECGs).
Select the patient record to be stored to the USB memory stick.
Select SYNC.