Product Manual
APPENDIX A
96
SIM Card Installation
Power off the ELI 150c. Open the printer cover and remove the pack of thermal paper. There is a small
access panel on the floor of the paper tray. Remove the screw and lift off the access panel.
Using your finger, slide the SIM card socket towards the back of the unit (follow the arrow printed on the
circuit board to the “open position”). Again, using your finger, lift the socket upright. To insert the SIM
card, hold the SIM card so the gold contact is facing you and the key (angle-notched corner) is in the
upper right.
Slide the SIM card in between the two recessed notches of the socket. Lower the socket so the SIM card key
appears in the lower right corner. Slide the socket towards the front of the unit (follow the arrow on the circuit
board to the “lock position”).
NOTE: Notched corner of SIM card must be located properly in socket. Do not force socket connector
if SIM is not seated properly.
Replace access panel, screw, and thermal paper. Power on the electrocardiograph.
If a wireless network carrier is not detected at the time of transmission (due to environmental conditions, i.e.,
poor signal), relocate or reposition the unit for a better signal and reattempt transmission.
In order to change a network carrier, you must scan the wireless network to determine what is detected and
available for use. Select F5 (Scan), a “scanning networks…” message will appear on the LCD. Once the
scanning process is completed, the Select Networks screen is available. Highlight the desired network and press
F3 (Selec).
Orders Download
NOTE: A custom ID must be downloaded before downloading the orders. Please reference the ELI
Link user manuals, and Custom ID Download in this section.
The device can download and process an ECG order list from ELI Link or another compatible electronic
information management system.
Order lists containing the demographic information of patients requiring an ECG are designed in ELI Link or an
E-S cribe system. The technician at the device selects the desired order code (e.g., a code specific to a
department or floor) and the patients belonging to the order list. Once downloaded to the device, the ECG list
for the selected order code is stored in the device as the order list (similar to the ECG directory). As with ECG
data transmission, you can use any of the connectivity options to download the order list.
From real-time ECG view, select F6 (More) followed by F3 (Orders Download). Use the appropriate
function key to select the desired study group.
Use F1 (▲) and F2 (▼) to scroll through the list; use F3 (Selec) to select the desired order code. Confirm or deny
your download by selecting F2 or F4.
“Transmission Status” will be displayed for approximately 10 seconds followed by “Dialing: telephone
number”, “Waiting for Response”, and “Connected”. Once connected, the screen indicates the number of
orders (ECGs) received for the order code. This only appears briefly before returning to the real-time ECG view.
When the order list has been downloaded, you may select the patients who need ECGs. Select F1 (ID) from
the real-time ECG view.