Product Manual

81
SECTION 4
Access Point Password
Access Point Password can be up to 120 characters in length.
Comm. Protocol
Select UNIPRO OR DICOM. DICOM is only available if the DICOM option has been installed.
NOTE: This parameter must be entered under the direction of the IT Manager of the facility where the
device is installed.
NOTE: Units ship by default with Comm. Protocol set to UNIPRO or DICOM. The UNIPRO or DICOM
setting is not supported by E-Scribe or ELI Link versions prior to V4.00. For questions about
compatibility of your device with ELI Link and UNIPRO or DICOM, contact Welch Allyn Technical
Support.
Sync Mode
Select None, XMT, or XMT+Orders (XMT+MWL if DICOM enabled). None requires a manual transmission of
reports and then a second manual request to receive orders from the cardiology management system. XMT
will automatically transmit the report; XMT+Orders will both transmit the report and retrieve the orders.
Sync Date/Time
Select Yes or No. Yes will synchronize the date/time with the approved cardiology management system. With
No, there will be no date/time synchronization. Date/time synchronization is done through ELI Link V4.00 or
later.
XMT Mandatory Fields
Defines fields required for ECG transmission to the cardiology management system. None will allow data
transmission without limitation; Last Name requires the technician to enter a minimum of the Last Name; ID
requires the technician to enter a minimum of the patient’s ID; Tech ID requires the technician to enter a
minimum of the technician’s initials. You may also select each of these independently allowing for any multiple
required selections as well.
Audit Trails
Enabling this option will create an audit log that records user actions and can be exported as a .TXT file via USB.
To manage audit logs:
1.
From real-time ECG view, select F6 (More).
2.
While holding down
(SHIFT)
, depress
ALT
and
D
simultaneously.
3. Select F4 (USB) to access the Software Upload/Download Screen where the audit trails are managed.
Selecting F4 in this screen will export the Audit Trails to a connected USB storage device.
Selecting F5 in this screen will delete the Audit Trails stored on the ELI 150c.
NOTE: Once maximum storage size is reached, the oldest audit trail file shall be deleted and a new
one shall be created to continue logging user actions.
Each audit log file begins with a date and time of creation, software version, cart number, and the device serial
number. Each entry in an audit trail file includes the date and time including time zone offset from UTC, user
name (if logged in), action name, what is affected by the action (if applicable) and associated data (if
applicable). The following actions and associated data are logged: