Product Manual
SECTION 4
74
If an unknown language is visible, use the following steps to revert to the language of your country:
1. F6 (More) from real-time ECG view.
2. Select F5 (Set Time/Date).
3. Simultaneously press
(SHIFT) + ALT + C.
4. Enter password (“admin”)
5. Press F2 (▼) four times.
6. Press F3 (►) until the desired language appears.
7. F6 (Exit) to return to real-time ECG view.
Alphabets of specific languages may require use of special characters in demographic fields. This is
accomplished by using the SYM key on the keyboard.
Volume
Defines the keyboard click loudness. Available settings range from 0 (off) to 8 (loud).
Battery Time Out
Determines when the electrocardiograph will switch off in order to conserve the battery life of the device. The
battery time out will only occur if the keyboard has not been depressed for the time specified. The battery time
out setting is ignored if an active ECG signal is detected during transmission or while rhythm printing.
ECG Storage
Indicates ECG storage capacity. Normal indicates standard memory capacity of 40 patient records. Expanded
indicates the optional expanded memory (200 patient records) has been installed.
ID Format
Defines the format for the patient demographic information prompts. There are two standard formats: short or
long. A custom ID format can be downloaded from ELI Link or an E-Scribe system. See Appendix A to download
a custom ID.
The short format includes the patient's last and first name, patient ID number, date of birth (automatically
calculates the age), and gender.
The long format is identical to the standard format except that it includes the patient's first name, room,
and comment fields.
Auto-Fill ID
When enabled, the device will automatically populate last name, first name, date of birth, age, and gender in
the ID screen if records with matching patient ID are found in the ECG directory.