Product Manual
65
SYSTEM SETTINGS
SECTION 4
Configuring Users and Roles
Options for configuring users and roles depends on whether Log-In Authentication is set to ON or OFF in
Configuration Settings. If Log-In Authentication is OFF, three generic roles, Technician, Site Admin, and
Admin, can be activated to manage access to specific device features.. If Log-In Authentication is ON, up to
30 unique users can be configured for the device. The default device setting has Log-In Authentication set
to OFF, and no active roles.
Configuring roles when Log-In Authentication is OFF
When Log-In Authentication is OFF, a user can access the basic ECG functionality of the ELI 150c without
entering a password. If the user attempts to access features that require additional authorization, they will
be prompted to input the password of an authorized role. The table below shows which features are
accessible by which roles. The Admin roll has access to all device features.
NOTE: If the password for the Technician remains blank, a guest to the system will have the same access as
described in the Technician column
To configure the roles:
1. From real-time ECG view, select F6 (More) followed by F5 (Set Time/Date).
2.
While holding down
•
(SHIFT)
, depress
ALT
and
P
simultaneously.
3. If required, enter password. This will automatically advance you to the set passwords display.
4. Select which role you want to set the password for: Technician, Site Admin, or Admin.
NOTE: The factory default password for the Admin is “admin” (lowercase, no quotation marks); the default
for the Site Admin is “siteadmin”; the default for the Technician is a blank field. It is suggested that the
password be changed after installation of the unit if utilizing role-based access features.
5. Enter a password for the selected role followed by a second entry to confirm.
NOTE: Password is case sensitive and alphanumeric.
6. From this display, select F6 (Exit) to return to real-time ECG view.