Product Manual

SECTION 2
54
Log-In Screen
When the Log-In Authentication mode is enabled (see Systems Settings section), and the device is powered on
or comes out of standby mode, a username/password prompt will display. If the Log-In Authentication is not
enabled (default setting), the device will proceed to the real-time ECG view.
To log-in, enter a username and password that match credentials found in the device’s user list (see
Configuration Settings section for details on configuring the user list and default password settings). Successful
log-in grants access based on role, which is set in the users list. Log off occurs after 10 minutes of inactivity.
Selecting “Guest” on the log-in screen bypasses the username and password entry. This allows quick access to
the ECG functionality and gives the ability to configure the user list..
Setting Time and Date
1. From real-time ECG view, select F6 (More) followed by F5 (Set Time/Date).
2. Use Enter, Tab, F1 (), or F2 () to move through each row. Use the keyboard to type in desired values
for the date and time (using a 24-hour clock).
NOTE: To set time immediately via auto synchronization press
F3 (Sync)
.
3. Use F3 () to move through the selections for setting Time Zone and Daylight Savings. To use Daylight
Savings, select Yes. Use F2 () to scroll to, or F4 (Page) to move to the begin/end settings page. Enter
the month, day, and time to begin Daylight Savings and the month, day, and time to end Daylight Savings.
Use F1 (), F2 (), or F4 (Page) to return to the previous screen. If the time zone selected does not
support
Daylight Savings, customize a begin and end time by selecting Custom. The Custom setting can also be used
to override the current Daylight Savings settings
TIP: Use the
BKSP
key to erase entry errors.
NOTE:
F4 (Page)
is only applicable to view read only (Yes) or change (Custom) a Daylight Savings
setting.
F4 (Page)
cannot be accessed from the Time Zone setting field.
4. Select F5 (Save) to save changes before exiting.
5. Select F6 (Exit) to return to real-time ECG view. If you did not save before selecting Exit, any changes
made to the time or date will be lost.
NOTE: Date and time may be set to auto synchronization with the cardiology management
system if available. (See section 4, Settings.)
NOTE: In the event of a hard reboot or the loss of battery power, the device will require the date and
time to be re-entered. The device will present a message stating “Set Date/Time.” The user then by
depressing any key (except alt, shift or power) will enter the date and time entry menu. You may bypass
this by selecting
F6 (Exit)
or
F5 (Save)
.
Important Version Information for WAM (Wireless Acquisition Module)
There are two generations of WAM (Wireless Acquisition Module) and UTK (USB Transceiver Key). A legacy WAM
and UTK and a newer version 2 WAM and UTK.