Product Manual
SECTION 1
44
ELI 150c Specifications
WAM / UTK
Radio specifications and certification information for the Wireless Acquisition Module (WAM) and USB Transceiver
Key (UTK), can be found in the WAM user manual.
Feature Specifications
Instrument Type 12-lead electrocardiograph
Input Channels Simultaneous acquisition of all 12 leads
Standard Leads Acquired I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
Waveform Display Backlit, ¼ VGA color LCD (320 x 240); 3,
4+4, or 6+6 lead presentation
Input Impedance Input
Dynamic Range
Electrode Offset Tolerance
Common Mode Rejection
Pacemaker Pulse Display
Meets or exceeds requirements of ANSI/AAMI/IEC 60601-2-25
Patient Leakage Current
Chassis Leakage Current
Meets or exceeds requirements of ANSI/AAMI ES1
Digital Sampling Rate 40,000 s/sec/channel used for pacemaker spike detection;
1000 s/sec/channel used for recording and analysis
Optional Functions Welch Allyn VERITAS resting ECG interpretation algorithm with age and gender
specific criteria; connectivity with bidirectional communication
Paper Perforated double Z-fold thermal paper; 108 mm (4”) wide, 200 sheets
Thermal Printer Computer-controlled dot array; 8 dots/mm
Thermal Printer Speeds 5, 10, 25, or 50 mm/s
Gain Settings 5, 10, or 20 mm/mV
Report Print Formats Standard or Cabrera; 3, 3+1, 3+3, or 6 channel
Rhythm Print Formats 3 or 6 channel with configurable lead groups
Keyboard Elastomer keyboard with complete alphanumeric keys, soft-key menu, and
dedicated function keys
Frequency Response 0.05 to 300 Hz
Filters High-performance baseline filter; AC interference filter 50/60 Hz; low-pass
filters 40 Hz, 150 Hz, or 300 Hz
A/D Conversion 20 bits (1.17 microvolt LSB)
Device Classification Class I, Type CF defibrillation-proof applied parts
ECG Storage
v1.x software - Normal -100 ECGs Expanded – 200 ECGs
v2.x software - Normal - 40 ECGs Expanded – 200 ECGs
Weight 7.2 lbs. (3.3 kg) including battery (without paper)
Dimensions 11.25 x 11.5 x 3.75” (29.2 x 30.5 x 10.2 cm)
Power Requirements Universal AC power supply (100-240 VAC at 50/60 Hz) 110 VA; internal
rechargeable battery
Battery Rechargeable sealed lead-acid (SLA) 12V battery; rated 2.2 Watts/Cell at 20
hours; 6.97” x 1.34” x 2.6” (177 x 34 x 66 mm); weighs 1.76 lbs. (0.80 kg)