Product Manual

USER SAFETY INFORMATION
12
This product complies with relevant electro-magnetic interference, mechanical safety, performance,
and biocompatibility standards. However, the product cannot completely eliminate potential patient or
user harm from the following:
Harm or device damage associated with electro-magnetic hazards,
Harm from mechanical hazards,
Harm from device, function, or parameter unavailability,
Harm from misuse error, such as inadequate cleaning, and/or
The device and IT Network the device is connected to should be securely configured and maintained
per the IEC 80001 standard, or an equivalent network security standard or practice.
Caution(s)
To prevent possible damage to the keyboard, do not use sharp or hard objects to depress keys,
only use fingertips.
Do not attempt to clean the device or patient cables by submersing into a liquid, autoclaving, or steam
cleaning as this may damage equipment or reduce its usable life. Wipe the exterior surfaces with a warm
water and mild detergent solution and then dry with a clean cloth. Use of unspecified cleaning/disinfecting
agents, failure to follow recommended procedures, or contact with unspecified materials could result in
increased risk of harm to users, patients and bystanders, or damage to the device.
No user-serviceable parts inside. Screw removal by qualified service personnel only. Damaged or
suspected inoperative equipment must be immediately removed from use and must be
checked/repaired by qualified service personnel prior to continued use.
The rechargeable internal battery is a sealed lead-acid type and it is totally maintenance free. If the
battery appears to become defective, refer to Welch Allyn Service Department.
Do not pull or stretch patient cables as this could result in mechanical and/or electrical failures. Patient
cables should be stored after forming them into a loose loop.
No calibration or special equipment is needed for the proper operation or maintenance of the device.
When necessary, dispose of the device, its components and accessories (e.g., batteries, cables,
electrodes), and/or packing materials in accordance with local regulations.
Use only No. 26 AWG or larger telecommunication line cord.
Note(s)
Patient movements may generate excessive noise that may affect the quality of the ECG traces and the
proper analysis performed by the device.
Proper patient preparation is important to proper application of ECG electrodes and operation of the device.
The algorithm detecting electrode misplacements is based on normal physiology and ECG lead order, and
tries to identify the most likely switch; however, it is advisable to check the other electrode positions in the
same group (limb or chest).
There is no known safety hazard if other equipment, such as pacemakers or other stimulators, is
used simultaneously with the device; however, disturbance to the signal may occur.