Product Manual

39
SECTION 4
Access Point Password
Access Point Password can be up to 120 characters in length.
Comm. Protocol
Select UNIPRO OR DICOM. DICOM is only available if the DICOM option has been installed.
NOTE: This parameter must be entered under the direction of the IT Manager of the facility where the
device is installed.
NOTE: Units ship by default with Comm. Protocol set to UNIPRO or DICOM. The UNIPRO or DICOM
setting is not supported by E-Scribe or ELI Link versions prior to V4.00. For questions about compatibility
of your device with ELI Link and UNIPRO or DICOM, contact Mortara Technical Support.
Sync Mode
Select None, XMT, or XMT+Orders (XMT+MWL if DICOM enabled). None requires a manual transmission of
reports and then a second manual request to receive orders from the cardiology management system. XMT will
automatically transmit the report; XMT+Orders will both transmit the report and retrieve the orders.
Sync Date/Time
Select Yes or No. Yes will synchronize the date/time with the approved cardiology management system. With No,
there will be no date/time synchronization. Date/time synchronization is done through ELI Link V4.00 or later.
XMT Mandatory Fields
Defines fields required for ECG transmission to the cardiology management system. None will allow data
transmission without limitation; Last Name requires the technician to enter a minimum of the Last Name; ID
requires the technician to enter a minimum of the patient’s ID; Tech ID requires the technician to enter a minimum
of the technician’s initials. You may also select each of these independently allowing for any multiple required
selections as well.