SECTION 2
Setting Time and Date
1. From real-time ECG view, select F6 (More) followed by F5 (Set Time/Date).
2. Use Enter, Tab, F1 (▲), or F2 (▼) to move through each row. Use the keyboard to type in desired values for
the date and time (using a 24-hour clock).
NOTE: To set time immediately via auto synchronization press F3 (Sync).
3. Use F3 (►) to move through the selections for setting Time Zone and Daylight Savings. To use Daylight
Savings, select Yes. Use F2 (▼) to scroll to, or F4 (Page) to move to the begin/end settings page. Enter the
month, day, and time to begin Daylight Savings and the month, day, and time to end Daylight Savings. Use
F1 (▲), F2 (▼), or F4 (Page) to return to the previous screen. If the time zone selected does not support
Daylight Savings, customize a begin and end time by selecting Custom. The Custom setting can also be used to
override the current Daylight Savings settings
TIP: Use the BKSP key to erase entry errors.
NOTE: F4 (Page) is only applicable to view read only (Yes) or change (Custom) a Daylight Savings
setting. F4 (Page) cannot be accessed from the Time Zone setting field.
4. Select F5 (Save) to save changes before exiting.
5. Select F6 (Exit) to return to real-time ECG view. If you did not save before selecting Exit, any changes made
to the time or date will be lost.
NOTE: Date and time may be set to auto synchronization with the cardiology management system if
available. (See section 4, Settings.)
NOTE: In the event of a hard reboot or the loss of battery power, the device will require the date and time
to be re-entered. The device will present a message stating “Set Date/Time.” The user then by depressing
any key (except alt, shift or power) will enter the date and time entry menu. You may bypass this by
selecting F6 (Exit) or F5 (Save).
Using the WAM or AM12 Acquisition Module
ECG acquisition and rhythm strip printing can also be performed at the WAM or the AM12 acquisition modules.
To use the WAM, refer to the WAM user manual. To use the AM12, refer to the AM12 short-form instruction card.
NOTE: The device must be configured at the factory for use with the WAM. Select F6 (More) followed by
F6 (More) to determine the device’s setting. “WAM Option Not Available” will display if the
device is not configured to work with the WAM.
NOTE: WAM must be paired to electrocardiograph before operation.
Installing the WLAN Antenna
The device with optional WLAN module is shipped with the antenna not installed: the antenna can be found in the
accessory box.
1. Remove the antenna from the accessory box.
2. Locate the antenna connector on the back of the device.
3. Mount the antenna on the connector by rotating the antenna clockwise. The antenna must be finger tight to its
connector.
4. Locate the built-in hinge and fold the antenna (it will now be at a 90° angle); continue to rotate the antenna
clockwise until it is placed vertically. This will guarantee the best signal for the WLAN module.
NOTE: For more information about the use of the WLAN option, refer to Appendix A.