ELI 150C 12-LEAD RESTING ECG
Feature Specification*
Instrument Type 12-lead electrocardiograph
Input Channels Simultaneous acquisition of all 12 leads
Standard Leads Acquired I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
Waveform Display Backlit, 1/4 VGA color LCD (320 x 240); 3, 4+4, or 6+6 lead presentation
Input Impedance
Input Dynamic Range
Electrode Offset Tolerance
Common Mode Rejection
Meets or exceeds the requirements of ANSI/AAMI EC11
Patient Leakage Current
Chassis Leakage Current
Meets or exceeds the requirements of ANSI/AAMI ES1
Digital Sampling Rate
40,000 s/sec/channel used for pacemaker spike detection;
1000 s/sec/channel used for recording and analysis
Optional Functions
VERITAS resting ECG interpretation algorithm with age and gender-specific criteria;
connectivity with bidirectional communication
Paper Perforated double Z-fold thermal paper; 108 mm (4”) wide, 200 sheets
Thermal Printer Computer-controlled dot array; 8 dots/mm
Thermal Printer Speeds 5, 10, 25, or 50 mm/s
Gain Settings 5, 10, or 20 mm/mV
Repot Print Formats Standard or Cabrera; 3, 3+1, 3+3, or 6 channel
Rhythm Print Formats 3 or 6 channel with configurable lead groups
Keyboard
Elastomer keyboard with complete alphanumeric keys, soft-key menu and dedicated function
keys
Frequency Response 0.05 to 300 Hz
Filters
High-performance baseline filter; AC interference filter 50/60 Hz; low-pass filters 40 Hz, 150 Hz,
or 300 Hz
A/D Conversion 20 bits (1.17 microvolt LSB)
Device Classification Class I, Type CF defibrillation-proof applied pats
ECG Storage Internal storage up to 40 ECGs; optional expanded up to 200 ECGs
Weight 7.2 lbs. (3.3 kg) including battery (without paper)
Dimensions 11.25 x 11.5 x 3.75” (29.2 x 30.5 x 10.2 cm)
Power Requirements Universal AC power supply (100-240 VAC at 50/60 Hz) 110 VA; internal rechargeable battery