Product Manual
Table Of Contents
- 1. Getting started
- 2. Recording an exercise ECG
- 2.1 The Real-time monitor
- 2.2 Record an exercise ECG
- 2.3 Real-time monitor
- 2.3.1. Select how you want to view leads
- 2.3.2. Measure blood pressure
- 2.3.3. Capture an event
- 2.3.4. Patient Perception Parameters
- 2.3.5. Duke Treadmill Score
- 2.3.6. Add a comment
- 2.3.7. Apply filters
- 2.3.8. Manual load change
- 2.3.9. Pacemaker pulse detection
- 2.3.10. Signal quality indication
- 2.3.11. Printing during recording
- 2.3.12. Target heart rate
- 2.3.13. Target Workload
- 3. Viewing an exercise ECG
- 3.1 About viewing exercise ECGs
- 3.2 View an exercise ECG
- 3.3 Work with the Summary view
- 3.4 Work with the Averages view
- 3.5 Work with the Full disclosure view
- 3.6 Work with the 2x6 and 4x3 view
- 3.7 Work with the Trends view
- 3.8 Work with the ST table view
- 3.9 Work with the Heart rate view
- 3.10 Work with the ST/HR view
- 4. Interpreting an exercise ECG
- 5. Printing an exercise ECG
- 6. Maintaining the recorder
- 7. Customizing the Exercise ECG module
- 8. Customize exercise protocols
- 9. Troubleshooting
- 10. Before installing the Exercise ECG software
- 11. Installing the Exercise ECG Module
- 12. Safety and precautions
- 13. Guidance and Manufacturer’s Declarations
- 14. Function keys
- 15. Additional information
CardioPerfect Workstation
Exercise ECG Module - User Manual
DIR 80012335 Ver. D 74 / 83
Electromagnetic Immunity
The Welch Allyn CardioPerfect Pro Cardiograph is intended for use in the electromagnetic environment specified
below. The customer or the user of the Welch Allyn CardioPerfect Pro Cardiograph should assure that it is used in
such an environment.
Immunity test IEC 60601
test level
Compliance level Electromagnetic environment
– guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output lines
±2 kV for power
supply lines
±1 kV for input/output
lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential
mode
±2 kV common mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input lines
IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 sec
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 sec
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the Welch Allyn CardioPerfect
Pro Cardiograph requires
continued operation during
power mains interruptions, it is
recommended that the Welch
Allyn CardioPerfect Pro
Cardiograph be powered from an
uninterruptible power supply or a
battery.
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
NOTE U
T
is the AC mains voltage prior to application of the test level.