Product Manual
Table Of Contents
- 1. Getting started
- 2. Recording an exercise ECG
- 2.1 The Real-time monitor
- 2.2 Record an exercise ECG
- 2.3 Real-time monitor
- 2.3.1. Select how you want to view leads
- 2.3.2. Measure blood pressure
- 2.3.3. Capture an event
- 2.3.4. Patient Perception Parameters
- 2.3.5. Duke Treadmill Score
- 2.3.6. Add a comment
- 2.3.7. Apply filters
- 2.3.8. Manual load change
- 2.3.9. Pacemaker pulse detection
- 2.3.10. Signal quality indication
- 2.3.11. Printing during recording
- 2.3.12. Target heart rate
- 2.3.13. Target Workload
- 3. Viewing an exercise ECG
- 3.1 About viewing exercise ECGs
- 3.2 View an exercise ECG
- 3.3 Work with the Summary view
- 3.4 Work with the Averages view
- 3.5 Work with the Full disclosure view
- 3.6 Work with the 2x6 and 4x3 view
- 3.7 Work with the Trends view
- 3.8 Work with the ST table view
- 3.9 Work with the Heart rate view
- 3.10 Work with the ST/HR view
- 4. Interpreting an exercise ECG
- 5. Printing an exercise ECG
- 6. Maintaining the recorder
- 7. Customizing the Exercise ECG module
- 8. Customize exercise protocols
- 9. Troubleshooting
- 10. Before installing the Exercise ECG software
- 11. Installing the Exercise ECG Module
- 12. Safety and precautions
- 13. Guidance and Manufacturer’s Declarations
- 14. Function keys
- 15. Additional information
CardioPerfect Workstation
Exercise ECG Module - User Manual
ï‚· Select a mode, see picture:
1. Select Timed mode to create a protocol which will run independent of the ergometer
or treadmill stages. An independent protocol consists of stages, each of which has its
own duration. Regardless of whether you hold or accelerate an exercise device
stage, the independent BP or print protocol will continue to run using the stage
intervals defined for that protocol.
2. If you select Follow Exercise protocol mode the Blood pressure action (prompt or
automatic BP start) or printout will occur at a fixed time after the start of each stage.
All stage related fields and menus will then be disabled. Any information in the pane
above the mode selector will be disregarded.
3. You can select a different mode (independent or Follow Exercise protocol) for each
phase
8.4 Protocol editor window
Use the Protocol editor window to create and edit protocols and their stages.
Protocol list
The left hand pane of the Protocol editor window contains a list with all existing protocols.
These protocols are divided into treadmill protocols, ergometer protocols, blood pressure
protocols and printer protocols.
Protocol workspace
The right hand pane of the Protocol editor window is where you add, edit and delete stages.
The workspace is divided into four tabs that represent the four phases in a protocol: pre-test,
exercise, recovery and post-test. For each phase, you can create a number of stages.
Menu bar
Here, you find functionality to import and export protocols, and to consult the help file.
DIR 80012335 Ver. D 60 / 83