Product Manual
Table Of Contents
- 1. Getting started
- 2. Recording an exercise ECG
- 2.1 The Real-time monitor
- 2.2 Record an exercise ECG
- 2.3 Real-time monitor
- 2.3.1. Select how you want to view leads
- 2.3.2. Measure blood pressure
- 2.3.3. Capture an event
- 2.3.4. Patient Perception Parameters
- 2.3.5. Duke Treadmill Score
- 2.3.6. Add a comment
- 2.3.7. Apply filters
- 2.3.8. Manual load change
- 2.3.9. Pacemaker pulse detection
- 2.3.10. Signal quality indication
- 2.3.11. Printing during recording
- 2.3.12. Target heart rate
- 2.3.13. Target Workload
- 3. Viewing an exercise ECG
- 3.1 About viewing exercise ECGs
- 3.2 View an exercise ECG
- 3.3 Work with the Summary view
- 3.4 Work with the Averages view
- 3.5 Work with the Full disclosure view
- 3.6 Work with the 2x6 and 4x3 view
- 3.7 Work with the Trends view
- 3.8 Work with the ST table view
- 3.9 Work with the Heart rate view
- 3.10 Work with the ST/HR view
- 4. Interpreting an exercise ECG
- 5. Printing an exercise ECG
- 6. Maintaining the recorder
- 7. Customizing the Exercise ECG module
- 8. Customize exercise protocols
- 9. Troubleshooting
- 10. Before installing the Exercise ECG software
- 11. Installing the Exercise ECG Module
- 12. Safety and precautions
- 13. Guidance and Manufacturer’s Declarations
- 14. Function keys
- 15. Additional information
CardioPerfect Workstation
Exercise ECG Module - User Manual
6. Maintaining the recorder
6.1 About maintaining the recorder
To keep your Welch Allyn CardioPerfect recorder in good working condition, please observe
the following:
Remove the battery when you do not use the recorder for a long period of time.
Do not expose the recorder to temperatures lower than -15 ºC or higher than 45 ºC,
at a maximum non-condensing humidity of 95%.
Clean the cables regularly.
Replace the battery of the recorder when the power indicator light is red.
6.2 Cleaning the Patient cables and PC interface cables
When you use electrodes that need electrode gel, make sure that the cables and the non-
disposable electrodes are cleaned regularly. Otherwise, the gel may build up on the wires.
Clean the cables and the non-disposable electrodes with a soft cloth moistened with a
recommended disinfectant or cleaning agent.
Cleaning cables: patient cables, plugs and power cords should be kept clean using lukewarm
soapy water or a neutral cleaner.
Disinfecting the cables: use chemical disinfectants containing ethanol (70%-80%), propanol
(70%-80%) or aldehydes (2%-4%).
Do not: Clean the patient cable with pure alcohol. Pure alcohol
can cause the plastic to become brittle and may cause
the cable to fail prematurely.
Caution
Do not: Autoclave the cable or use ultrasonic cleaners.
Do not: Immerse the
patient cable.
Do not:
Wet the c
onnectors.
6.3 Replace the battery of the recorder
6.3.1. Welch Allyn CardioPerfect MD recorders
The Welch Allyn CardioPerfect MD recorders get their power from a 9V alkaline battery or
optionally from an AC-mains adapter. If a 9V battery is used, you need to replace it when the
power indicator light becomes red instead of green. Do not use the recorder when the power
indicator led is red.