Product Manual
Table Of Contents
- 1. Getting started
- 2. Recording an exercise ECG
- 2.1 The Real-time monitor
- 2.2 Record an exercise ECG
- 2.3 Real-time monitor
- 2.3.1. Select how you want to view leads
- 2.3.2. Measure blood pressure
- 2.3.3. Capture an event
- 2.3.4. Patient Perception Parameters
- 2.3.5. Duke Treadmill Score
- 2.3.6. Add a comment
- 2.3.7. Apply filters
- 2.3.8. Manual load change
- 2.3.9. Pacemaker pulse detection
- 2.3.10. Signal quality indication
- 2.3.11. Printing during recording
- 2.3.12. Target heart rate
- 2.3.13. Target Workload
- 3. Viewing an exercise ECG
- 3.1 About viewing exercise ECGs
- 3.2 View an exercise ECG
- 3.3 Work with the Summary view
- 3.4 Work with the Averages view
- 3.5 Work with the Full disclosure view
- 3.6 Work with the 2x6 and 4x3 view
- 3.7 Work with the Trends view
- 3.8 Work with the ST table view
- 3.9 Work with the Heart rate view
- 3.10 Work with the ST/HR view
- 4. Interpreting an exercise ECG
- 5. Printing an exercise ECG
- 6. Maintaining the recorder
- 7. Customizing the Exercise ECG module
- 8. Customize exercise protocols
- 9. Troubleshooting
- 10. Before installing the Exercise ECG software
- 11. Installing the Exercise ECG Module
- 12. Safety and precautions
- 13. Guidance and Manufacturer’s Declarations
- 14. Function keys
- 15. Additional information
CardioPerfect Workstation
Exercise ECG Module - User Manual
2.3.13. Target Workload
If an Ergometer was selected as exercise device, the Real-time monitor displays the current
Workload and the current Workload as a percentage of the Target Workload during recording,
in the overview area.
The Target Workload can be adjusted at the start of a new exercise ECG in the New Exercise
ECG window.
The Standard Target Workload shown is calculated with the following formula:
Target Workload = ((2 * Height) - (2.37 * Age) - 73) * 1.01 + 16.6
In case of a female patient multiply the above by 0.8.
Units:
The height is specified in centimeters.
The Age is specified in years.
To set the Target Workload:
Enter a value in the Target Workload field.
DIR 80012335 Ver. D 27 / 83