Message Possible cause Suggested action Alarm priority
Check cuff inflation
settings.
Cuff inflation target was
overridden due to max pressure
being too low
Change cuff inflation target or
maximum pressure so that the cuff
inflation target is at least 20 mmHg
lower than maximum pressure.
Information
Tube type does not
match device
configuration.
Switching to step BP Change tube type to dual lumen or
change the algorithm configuration
to step BP
Information
NIBP not functional.
05FF01
Unrecognized WACP parameter
received from sensor
Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
05FF02
Timeout waiting for sensor
response
Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
05FF03
Error deserializing the WACP
message received from the sensor
Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
05FF04
WACP stack send message failure Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
05FF05
Timeout waiting for asynchronous
sensor message
Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
05FF06
One or more undetermined
numerics when reading status
indicates OK
Check connections. Restrict patient
movement.
Very low
NIBP not functional.
05FF07
Unrecognized sensor reading
status code
Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
05FF08
Sensor power up failure Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
05FF09
WACP rendezvous failure Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
05FF0A
Application firmware retrieval
error during POST
Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
05FF0B
Upgrade .pim file is corrupt Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
05FF0C
Configured upgrade firmware
directory cannot be accessed
Internal malfunction. If the problem
persists, replace the module.
Very low
Device configuration
issue. 05FF0D
Configured parameter (NIBP or
SpO2) used in Intervals is missing
Use configured parameters for
Intervals
Very low
NIBP not functional.
05FF0E
The NIBP sensor unexpectedly
reset
Clear the error and try again Very low
NIBP not functional.
05FF0F
The NIBP sensor firmware failed
to upgrade
Internal malfunction. If the problem
persists, replace the module.
Very low
92 Troubleshooting Connex
®