Product Manual
Message Possible cause Suggested action Alarm priority
Stat mode requested too
soon. Clear to retry.
050609
NIBP excessive stat mode time.
The time between readings is less
than one minute and the readings
plus time between readings result
in the device taking over 15
minutes to complete the averaging
cycle.
Internal malfunction. If the problem
persists, replace the module.
Very low
Unable to determine
NIBP; check connections
and tubing for kinks.
05060A
NIBP transducers are not matched The transducer are above 5 mmHg
and the difference in pressure is
greater than 40 mmHg. Check cuff
for pinched or obstructed tubing. If
the problem persists, replace the
module.
Medium
NIBP not calibrated.
Calibrate the module.
05060B
Factory EEPROM checksum error
on NIBP. Units internal
configuration was corrupted
Calibrate the NIBP module. If the
problem persists, replace the
module.
Very low
NIBP not functional.
05060C
NIBP command not implemented Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
05060D
NIBP wrong data count Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
05060E
NIBP data range error Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
05060F
NIBP no POST error to clear Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
050610
NIBP cannot clear this POST error Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
050611
NIBP command not command type Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
050612
NIBP communication timeout Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
050613
NIBP response header wrong Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
050614
NIBP response checksum wrong Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
050615
Too much NIBP data was received Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
050616
NIBP FPROM erase error Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
050617
NIBP FPROM programming error Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
050618
Invalid NIBP target pressure Internal malfunction. If the problem
persists, replace the module.
Very low
Directions for use Troubleshooting 91