Product Manual

Message Possible cause Suggested action Alarm priority
Invalid algorithm. Select
correct algorithm and
retry. 05050A
Invalid NIBP Algorithm. NIBP
component software tried to
configure the sensor in an illegal
manner
Verify the algorithm. If the problem
persists, replace the module.
Very low
NIBP not functional.
050513
Invalid NIBP initiation code Internal malfunction. If the problem
persists, replace the module.
Very low
Invalid patient mode.
Select correct patient
mode and retry. 050514
Invalid patient mode on NIBP. The
NIBP component software tried to
configure the sensor in an illegal
manner
Verify correct patient mode. If the
problem persists, replace the
module.
Very low
NIBP not functional.
050515
Invalid module configuration for
NIBP
Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
050516
NIBP module malfunction Internal malfunction. If the problem
persists, replace the module.
Very low
Ambient temperature
out of range. Clear error
and retry. 050517
Ambient temp out of range on
NIBP
Restore device to normal
temperature ranges and try again.
Very low
Low battery. Plug into
outlet. 050518
The NIBP module power rail is too
low
Plug device into an AC outlet to
charge the battery.
Very low
Battery overcharged.
Disconnect from outlet.
050519
The NIBP module power rail is too
high.
Battery is overcharged. Remove from
charging source.
Very low
NIBP not calibrated.
Calibrate the module.
050601
NIBP failed to load safety
processors calibration record from
EEPROM
Calibrate the NIBP module. If the
problem persists, replace the
module.
Very low
NIBP not functional.
050602
NIBP safety processor failed ROM
checksum
Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not calibrated.
Calibrate the module.
050603
NIBP safety processor not
calibrated, missing calibration
signature
Calibrate the NIBP module. If the
problem persists, replace the
module.
Very low
Cuff pressure limits
exceeded. 050604
NIBP system failure. Overpressure Restrict patient movement. Medium
Premature auto cycle
skipped. 050605
NIBP auto cycle skipped, SVRP
requirement not satisfied
Cuff pressure is not below safety
return pressure long enough to allow
a cycle to occur.
Very low
Cuff pressure too high.
Clear error to retry.
050606
NIBP cuff pressure above SVRP for
to long
Verify cuff connections. If the
problem persists, replace the
module.
Medium
NIBP not functional.
050607
NIBP cannot clear the failsafe
alerts
Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
050608
NIBP safety processor has stopped
responding
Internal malfunction. If the problem
persists, replace the module.
Very low
90 Troubleshooting Connex
®
Spot Monitor