Product Manual
Message Possible cause Suggested action Alarm priority
patient movement.
05000F
NIBP not functional.
050105
WACP message CRC mismatch on
NIBP module
Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
050201
This message is not implemented
by the NIBP module
Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
050202
This message is not supported by
the NIBP module
Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
050203
The NIBP module has run out of
memory
Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
050205
The NIBP module has received a
invalid parameter
Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
050206
The parameter provided by the
NIBP module is outside of the
allowable range for the specified
message
Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
050207
The NIBP module message
requires an object, but did not
contain one
Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
050208
The NIBP module object provided
with the message could not be
deserialized
Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
050209
The NIBP module object could not
be serialized
Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
05020A
The NIBP module message is
performing a request or action
when the module state prohibits
the request or action.
Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not calibrated.
050503
Factory EEPROM checksum error
on NIBP. Units internal
configuration was corrupted
Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not functional.
050504
User EEPROM checksum error.
Configuration data which can be
set in the user's configuration
menu was damaged or lost on
NIBP
Calibrate the NIBP Module. If
problem persists, replace module.
Very low
NIBP not functional.
050505
Post failure of A/D convertor Internal malfunction. If the problem
persists, replace the module.
Very low
NIBP not calibrated.
Calibrate the module.
050509
NIBP module calibration failure,
the calibration signature is zero
Calibrate the NIBP module. Very low
Directions for use Troubleshooting 89