Message Possible cause Suggested action Alarm priority
Unable to determine
NIBP; check
connections; limit
patient movement.
050003
The NIBP measurement may be
inaccurate, patient motion
occurred, or the settings for
patient readings obtain might not
be accurate
Make sure the NIBP settings/patient
mode is appropriate. If the problem
persists, replace the module.
Medium
Unable to determine
NIBP; check
connections; limit
patient movement.
050004
Excessive artifact, no blood
pressure parameters could be
calculated
Unable to determine blood pressure.
Check connections; restrict patient
movement. If problem persists,
replace module
Medium
Unable to determine
NIBP; check inflation
settings. 050005
Low inflation on blood pressure
measurement attempt
Make sure the NIBP settings/patient
mode is appropriate. If the problem
persists, replace the module.
Medium
Unable to determine
NIBP; check connections
and tubing for kinks.
050006
The NIBP tubing has a kink, or
there is a NIBP transducer
calibration failure
Internal malfunction. If the problem
persists, replace the module.
Medium
Unable to determine
NIBP; check
connections; limit
patient movement.
050007
The blood pressure measurement
dumped too soon
Make sure the NIBP settings/patient
mode is appropriate. If the problem
persists, replace the module.
Medium
Unable to determine
NIBP; check
connections; limit
patient movement.
050008
There were not enough steps on
the measurement attempt
Unable to determine blood pressure.
Check connections; restrict patient
movement.
Medium
Unable to determine
NIBP; check inflation
settings. 050009
There is invalid patient
information for the selected mode
Make sure the NIBP settings/patient
mode is appropriate. If the problem
persists, replace the module.
Medium
Unable to determine
NIBP; check
connections; limit
patient movement.
05000A
The re-inflation was too late on
the measurement attempt
Unable to determine blood pressure.
Check connections; restrict patient
movement.
Medium
Unable to determine
NIBP; check inflation
settings. 05000B
There were many re-inflate
attempts on the measurement
attempt
Unable to determine blood pressure.
Check connections; restrict patient
movement.
Medium
Unable to determine
NIBP; check connections
and tubing for kinks.
05000C
Couldn’t bleed pressure below
safe venous return pressure
Unable to release cuff pressure.
Check tubing for kinks and
connection integrity.
Medium
No display The safety check failed on the
measurement attempt
Internal malfunction. If the problem
persists, replace the module.
Unable to determine
NIBP; check
connections; limit
The NIBP pressure is not stable
and the transducer zero value
cannot be set
The NIBP pressure is not stable and
the transducer zero value cannot be
set. If problem persists, replace
module.
Medium
88 Troubleshooting Connex
®
Spot Monitor