Product Manual

2. Remove the thermometer from the dock and clean the metal electrical contacts on
the thermometer with the cotton swab.
3. Place the thermometer aside for 1 minute, allowing the contacts to air dry.
4. Clean the metal electrical contacts on the device dock with the cotton swab.
5. Allow the contacts to air dry for 1 minute.
6. Optional step to improve performance: On the dock only, apply DeoxIT GoldĀ® to
the tip of each electrical contact. Do not apply DeoxIT Gold to the thermometer.
No drying time is required.
7. Return the Braun thermometer to the dock.
Directions for use Maintenance and service 85