Product Manual
WARNING Electric shock hazard. Do not open the monitor or attempt
repairs. The monitor has no user-serviceable internal parts. Only perform
routine cleaning and maintenance procedures specifically described in this
manual. Inspection and servicing of internal parts shall only be performed
by qualified service personnel.
CAUTION Sterlizing the monitor could damage the device.
1. Disconnect the AC power cord from the mains outlet.
2. Wipe the top of the monitor.
3. Wipe the sides, front, and rear of the monitor.
4. Avoid residual film buildup on the LCD screen. After cleaning and disinfecting, wipe
the LCD screen with a clean cloth dampened with water, and then wipe the screen
dry with a dry clean cloth.
5. Wipe the bottom of monitor.
6. Wipe the APM work surface.
7. Wipe the accessory bins or basket.
8. Wipe the AC power cord and the APM work surface power/USB cable assembly.
9. Wipe the stand from top to bottom.
Dry the equipment
1. Allow all components except the LCD screen to air dry.
2. Wipe the LCD screen dry with a clean cloth.
Store the device
Store the device according to facility guidelines to keep the device clean, dry, and
ready for service.
Cleaning accessories
Accessories include components like blood pressure cuffs and hoses, SpO2 sensors and
cables, thermometers, and the barcode scanner. Follow accessory manufacturer’s
instructions for cleaning and disinfection.
For the Braun ThermoScan PRO 6000 thermometer, use only the approved cleaning
agents published in the manufacturer's instructions for cleaning. Unapproved
cleaning agents can damage the device and interfere with data transmission.
Clean the Braun ThermoScan PRO 6000 contacts
Debris that accumulates on the Braun ThermoScan PRO 6000 electrical contacts can
interfere with data transmission. Welch Allyn recommends cleaning the contacts on the
thermometer and the dock once every 4 months to maintain optimal performance.
CAUTION Do not use bleach solutions of any kind when cleaning metal
electrical contacts. They will damage the device.
1. Slightly dampen a cotton swab with 70% isopropyl alcohol.
84 Maintenance and service Connex
®
Spot Monitor