Product Manual
Section 2. Not approved for all Connex Spot Monitor components
Note The following cleaning agents are NOT approved for cleaning Connex Spot
Monitors equipped with the Braun ThermoScan PRO 6000
Cleaning agent Additional information
Cleancide
Clorox HealthCare Bleach Germicidal Cleaner
Sani-Cloth
®
Bleach
Not approved for cleaning the display
Super HDQ
®
L10
Dilution rate of ½ oz per gallon of water (1:256) applied to a
clean cloth
Tuffie5 Cleaning Wipes
Virex II (256) Dilution rate of ½ oz per gallon of water (1:256) applied to a
clean cloth
10 percent bleach solution (.5% - 1% sodium hypochlorite) applied to a clean cloth
Remove liquid spills from the monitor
Liquids can damage electronics inside the monitor. Follow these steps if liquids spill on
the monitor.
1.
Power down the monitor.
2. Disconnect the power cord from the mains outlet and the power source.
3. Remove battery pack from the monitor.
4. Dry excess liquid from the monitor.
5. Reinstall battery pack.
6. Reconnect the power cord.
7. Power on the monitor and verify that the monitor functions normally before using it.
If liquids possibly entered the monitor, remove the monitor from use until it has been
properly dried, inspected, and tested by qualified service personnel.
Clean the equipment
The screen lock blocks the display of patient information and prevents any input, which
may be useful when cleaning the display.
Follow the cleaning agent manufacturer's instructions to prepare solution, if applicable,
and clean all exposed surfaces of the monitor, APM work surface, accessory bin(s) and
basket, cords and cables, and stand. Wipe all surfaces until no visible soil remains.
Change the wipe or cloth throughout the cleaning procedure as needed.