CAUTION Do not sterilize the monitor. Sterilizing the monitor could harm
the device.
If liquids are spilled on the monitor:
1. Power down the monitor.
2. Disconnect the power cord from the mains outlet and the power source.
3. Remove battery pack from the monitor.
4. Dry off excess liquid from the monitor.
Note If liquids possibly entered the monitor, remove the monitor
from use until it has been properly dried, inspected, and
tested by qualified service personnel.
5. Reinstall battery pack.
6. Reconnect the power cord.
7. Power on the monitor and verify that the monitor functions normally before using it.
Prepare to clean the equipment
CAUTION Some cleaning agents are not appropriate for all components of
the device. Use only approved cleaning agents, and observe restrictions
noted for some components in the following table. Using unapproved
cleaning agents may cause damage to components.
CAUTION Do not use bleach solutions of any kind when cleaning metal
electrical contacts. They will damage the device.
Select a cleaning agent from the following table.
Section 1. Approved for all Connex Spot Monitor components
Cleaning agent
Additional information
Accel INTERVention
Accel TB
CaviWipes
Clinell® Universal Wipes
Oxiver TB
Sani-Cloth
®
Plus
Super Sani-Cloth
®
70 percent isopropyl alcohol solution Applied to a clean cloth
82 Maintenance and service Connex
®