3. Gently lift the latch with one hand and pull the tab on the top of the battery with your
other hand to remove the battery from its slot.
4. Slide the new battery into the slot.
Note Ensure that the tab is facing you on the top of the battery.
5. Replace the battery cover and tighten the captive screw on the bottom of the APM
work surface.
Cleaning requirements
This section presents procedures for cleaning the Connex Spot Monitor (including the
monitor, stand, APM work surface, accessories, and accessory basket and bins).
Welch Allyn has validated these instructions to be capable of preparing your Connex
Spot Monitor devices and above accessories for re-use. Clean on a routine basis
according to your facility's protocols and standards or local regulations. If the monitor is
on, lock the display.
WARNING Patient injury risk. Clean all accessories, including cables and
tubes, before storing the accessories on the device or stand. This helps
reduce the risk of cross contamination and nosocomial infection.
WARNING Electric shock hazard. Before cleaning the monitor, disconnect
the AC power cord from the mains outlet and the power source.
WARNING Electric shock hazard. DO NOT immerse or autoclave the
monitor or accessories. The monitor and the accessories are not heat-
resistant.
WARNING Liquids can damage electronics inside the monitor. Prevent
liquids from spilling on the monitor.
Directions for use Maintenance and service 81