Product Manual

4. Remove the old battery from the battery compartment.
5. Disconnect the battery connector from the battery connection port on the monitor.
6. Insert the battery connector for the new battery into the battery connection port on
the monitor.
7. Insert the new battery into the battery compartment.
8. Replace the battery cover, and then tighten the captive screw at the bottom of the
battery cover.
Note Do not overtighten the screw.
Replace the APM work surface battery
Before removing the APM work surface battery, power down the monitor and
disconnect the power cord from the mains outlet.
Note You do not have to remove the APM work surface from the stand to
remove the APM work surface battery.
1. Loosen the captive screw on the bottom of the APM work surface that secures the
battery cover.
2. Remove the battery cover and put it aside.
80 Maintenance and service Connex
®
Spot Monitor