Product Manual

WARNING Patient harm risk. Do not use intervals on neonates out of your
hearing range. Verify that the audio can be heard from where you intend to
be.
Automatic intervals
You can configure the monitor to take automatic NIBP measurements at consistent
intervals.
Note An alarm does not turn off intervals. Subsequent automatic measurements
continue to occur as scheduled.
Start Automatic intervals
1. Place the proper cuff around the patient’s bare upper arm.
2. On the Home tab, touch .
The vertical Intervals tab on the Settings tab appears.
3. Select Automatic.
4. Use the keypad to enter the length of time between NIBP measurements.
5. Touch Start intervals.
Program intervals
The monitor comes with six custom programs. One program is always available for you
to customize to meet your specific needs. If your facility does not configure all of the
remaining five programs, you can customize the remaining programs at any time.
The numbers below the program names indicate the length of time between each
interval in the cycle.
Start program intervals
You must be in either the Intervals or Office profile to access intervals.
1.
Place the proper cuff around the patient’s bare upper arm.
2. On the Home tab, touch .
The vertical Intervals tab on the Settings tab appears.
3. Select Program.
The preconfigured Program screen appears with the available programs and the
interval between measurements displayed to the right of the program.
4. Touch the program you want to use.
5. If you want to change the interval for the program selected, use the keypad to the
right of the program to enter the new interval.
6. Touch Start intervals.
Directions for use Patient monitoring 57