Alarm messages Priority
Alarm limit exceeded. SpO2 LOW. High
Alarm limit exceeded. Temperature HIGH. High
Alarm limit exceeded. Temperature LOW. High
Nurse Call
The monitor can be connected to a Nurse Call system through a cable that connects to
the Nurse Call connector.
When the Nurse Call cable is connected and Nurse Call is enabled, the monitor
immediately notifies the Nurse Call system when any alarm condition occurs that
exceeds the preset threshold. The Nurse Call system is also synchronized with the alarm
frame and audible alerts on the monitor.
Nurse Call thresholds are set in the configuration settings.
To connect the monitor to a Nurse Call system, you must have a cable that has been
adapted to your Nurse Call system (REF 6000-NC), rated 25V AC or 60V DC maximum at
1A maximum. For ordering information, see Approved Accessories in the Appendix.
WARNING Do not rely exclusively on Nurse Call for patient monitoring.
Although the Nurse Call option enables remote notification of an alarm
condition, it is not intended to replace appropriate bedside patient
monitoring by trained clinicians.
Note When a patient alarm occurs, touching the alarm icon in the Device Status
area pauses the alarm tone for 1 minute, as specified in the default settings
in Advanced Settings, but the visual alarm indicator(s) on the monitor and
Nurse Call continue.
Directions for use Alarms 53