Product Manual

Reset (pause or turn off) audio alarms
Audio alarm characteristics
After you reset an audio alarm, some tones do not return, but others return after a
pause interval if the condition that caused the alarm persists. Settings in the
Advanced tab determine the length of the pause interval.
If a new alarm condition occurs during a pause interval, a new audio tone occurs.
Pause or turn off an audio alarm
1. In the Device Status area, touch .
Visual indications remain in the parameter frame until the condition is corrected
or until the next measurement is taken.
In the Device Status area, if the icon changes to and the message
remains, the timer counts down and the audio tone returns after a pause
interval. You can touch
again to restart the timer.
If you responded to an NIBP alarm and multiple NIBP limits have been
exceeded, the first audio tone and message go away, but another NIBP limit
message shows with a countdown timer. A new NIBP audio tone sounds after
the countdown unless you touch to dismiss each remaining NIBP limit
message.
2. If multiple alarms are active, a multiple alarm toggle will appear in the Device Status
area. Respond to multiple alarms as follows:
a. Touch in the Device Status area. (See note below.)
b. Read the alarm message for the second alarm.
c. Touch .
d. Continue to touch multiple alarm toggle buttons and to reset tones until you
have read all of the messages.
Note The multiple alarm toggle button will display the number
of active alarms inside the alarm icon. A set of dots
indicating the display order of alarms from highest (left)
to lowest (right) priority (as well as the most recent in
the case of multiple alarms of the same priority) will
appear below it.